Endoscopy AccuTouch® System FAQQ: When stepping on the foot pedal I get no response in the simulation. Q: During EGD the scope tube seems to pushing back as I try to start the simulation. Q: When using the snare in a procedure I see no tool. Q: Simulation freezes or becomes erratic. Q: I get the message "can't connect to device." Q: When stepping on the foot pedal I get no response in the simulation.A: Foot pedal is plugged into the wrong jack. For a single foot pedal plug into the lower jack. For a dual foot pedal the left foot pedal plugs into the top jack and the right foot pedal plugs into the lower jack. Do not use the middle jack.
Q: During EGD the scope tube seems to pushing back as I try to start the simulation.A: Anatomy face plate is misaligned. Try removing and reinstalling the face plate.
Q: When using the snare in a procedure I see no tool.A: No tool was selected. Go to accessories and select a tool.
Q: Simulation freezes or becomes erratic.A: Snare is plugged in but not completely seated. Either remove the snare or plug it in completely.
Q: I get the message "can't connect to device."A: Manikin is powered off or the USB cable is not connected to the PC
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||
|